Short Communication

Detection limit of a lutetium based non-paralizable PET-like detector

Emmanuel Busato* and Édouard Roux

Published: 16 April, 2020 | Volume 3 - Issue 1 | Pages: 063-069

The effect of the intrinsic lutetium radioactivity on the detection performances of a LYSO based in-beam PET-like prototype used for quality control of hadrontherapy treatments is studied. This radioactivity leads to a background that degrades the measurement of the β+ signal. In particular, it prevents the measurement of faint signals originating from low activity β+ sources. This paper presents a method to estimate the minimum β+ activity that can be measured for any acquisition time taking into account the non-extensible dead time of the detector. This method is illustrated with experimental data collected with the in-beam PET-like prototype. The results presented in this paper are therefore specific to this detector. The method can however be applied in other contexts, either to other lutetium based PET detectors or even to non-PET detectors affected by lutetium radioactivity. The dead time correction formalism can also be used generally to scale signal and background yields in any non-paralizable detector, even those in which the background is not due to the presence of intrinsic radioactivity.

Read Full Article HTML DOI: 10.29328/journal.ijpra.1001024 Cite this Article Read Full Article PDF

References

  1. Binet S, Bongrand A, Busato E, Force P, Guicheney C, et al. Construction and first tests of an in-beam PET demonstrator dedicated to the ballistic control of hadrontherapy treatments with 65 MeV protons. IEEE Transactions on Radiation and Plasma Medical Sciences. 2018; 2: 51–60.
  2. Kossert K, Jörg G, Gostomski CL. Experimental half-life determination of 176Lu. Applied Radiation and Isotopes, 81:140 – 145, 2013. 6th International Conference on Radionuclide Metrology - Low Level Radioactivity Measurement Techniques.
  3. Knoll GF. Radiation detection and measurement. 4th ed. Wiley, New York, NY, 2010.
  4. Choudalakis G, Casadei D. Plotting the differences between data and expectation. Eur Physical J Plus. 2012; 127: 25.
  5. Antcheva I, Ballintijn M, Bellenot B, Biskup M, Brun R, et al. ROOT — a C++ framework for petabyte data storage, statistical analysis and visualization. Computer Physics Communications. 40 YEARS OF CPC: A celebratory issue focused on quality software for high performance, grid and novel computing architectures. 2009 180: 2499 – 2512.
  6. Busato E, Calvet D, Theveneaux-Pelzer T. OPTHYLIC: An Optimised Tool for HYbrid Limits Computation. Computer Physics Communications. 2018; 226: 136 – 150.
  7. Parodi K, Enghardt W. Potential application of PET in quality assurance of proton therapy. Physics in Medicine & Biology. 2000; 45: N151.

Figures:

Figure 1

Figure 1

Figure 1

Figure 2

Figure 1

Figure 3

Figure 1

Figure 4

Figure 1

Figure 5

Figure 1

Figure 6

Similar Articles

Recently Viewed

Read More

Most Viewed

Read More

Help ?